FYI NOTE
FYI010598
Reversed Phase and
Ion Exchange Chromatography of Two Proteins on STYROS™
Columns.
|

|
Operating
parameters:
|
HPLC System. |
HP 1100, standard cell. |
|
Column |
STYROS™ 2R/XH
50 x 4.6mm |
|
Mobile Phase |
A: 0.1% TFA in H2O
B: 0.1% TFA in Acetonitrile/H2O (95/5)(v/v) |
|
Flow rate |
2.5 ml/min (900cm/hr) |
|
Gradient |
15 to 80% B in 6.25 Column Volume |
|
Temperature |
Ambient |
|
Detection |
214 nm |
|
Injection volume |
20 ml
|
|
Sample
(1 mg/ml each) |
1- 99% pure BSA.
2- 99% pure Ovalbumin |
|
|

|
Operating
parameters:
|
HPLC System. |
HP 1100, standard cell. |
|
Column |
STYROS™ Q
/NB
250 x 2.1mm |
|
Mobile Phase |
A: 20 mM Tris , pH= 8.2
B: A + 1 M NaCl, pH= 8.2 |
|
Flow rate |
2 ml/min (3,470 cm/hr) |
|
Gradient |
10% B to 40% B in 31 Column Volume |
|
Temperature |
Ambient |
|
Detection |
280 nm |
|
Injection volume |
5 ml
|
|
Sample
(5 mg/ml each) |
1,2,3,4,5-: Components of a 99% pure sample of
Ovalbumin.
6,7- : Components of a 99 % pure sample of BSA. |
|
|